Identifiler Direct

Validated Protocol for Processing with Identifiler® Direct

The following protocol has been validated for processing the Bode Buccal DNA Collector with the AmpFlSTR® Identifiler® Direct PCR Amplification Kit.

Remove one 1.2mm punch from the Buccal DNA Collector and place into tube or PCR plate well.
Add Bode PunchPrep Solution (Cat No. P09D79) and incubate on a 70oC heat block for 20 minutes.
Add 25µl of AmpFlSTR® Identifiler® Direct reaction mix to the 1.2 mm punch and perform a *26-cycle PCR reaction according to the instructions found in the AmpFlSTR® Identifiler® Direct PCR Amplifcation Kit User Guide.

*It is recommended that internal validation and testing regarding the number of PCR cycles is performed by each agency. The number of cycles within this protocol should serve as a recommendation for processing.


Validation Results

Bode Buccal DNA Collector FTA Sample
Number of Samples 88 88
1st Pass Success Rate 84% 82%

The Buccal DNA Collector provided equivalent performance to FTA when processing paired samples using the AmpFlSTR® Identifiler® Direct PCR Amplification Kit.

Electropherogram 1: Sample from the Buccal DNA Collector – processed with AmpFlSTR® Identifiler® Direct PCR Amplification Kit

Electropherogram 2: Sample from the same donor collected on FTA paper – processed with AmpFlSTR® Identifiler® Direct PCR